게시판
한외여과장치 5set 구매공급사업 견적요청의 E-Market 메시지
1식 ₩100,500,000원 이하 120일 이내 충청권에 설치도
검토와 일괄총액의 견적서로 답신주시고 매출증대일익을 협의하십시요
누락파일이나 추가적인 자료요청은 통화대신 메일소통이 효율적입니다
성과를 위한 관급프로젝트의 파트너사 - 제이앤에스 관리자와 운영인
관리인 메일 nhbbc@naver.com sms 070-4644-8617 팩스 050-5115-5372
운영자와의 멘토 010-5670-5227 보이스마켓과 쇼핑몰은 정중히 사양
I. 용도(End-user's Use)
1. 단백질 농축 및 투석 시 사용 될 장비
2. Pyrogen, NaCl, Salt 제거 및 Buffer Exchange에 용이한 장비
3. 세포배양 및 바이러스배양 정제시 필요한 장비
II. 장비의 구성(System Configuration of Goods)
Part I
1. 본체 (Main body)
1) Ultrafiltration/Diafiltration Main Unit 4set
2. Accessories
1) UF membrane 50cm2 12pk
2) Multi Manifolder 4pk
3) One Way Stop Cock Valve 1pk
4) Tank vent filter 1pk
Part II
1. 본체 (Main body)
1) Ultrafiltration/Diafiltration Main Unit 1set
2. Accessories
1) UF membrane 0.1m2 3pk
2) Retentate Flow Meter kit 1pk
3) Transfer pump console(230VAC) 1pk
III. 성능 및 규격(Performance and Specification)
A. FEATURE
Part I
1. Applications
1) Concentration and diafiltration
2) Depyrogenation
3) Desalting and buffer exchange
4) Small molecule processing
5) Cell harvesting/ clarification
6) Virus harvesting/ clarification
2. Feature of System
1) Single unit system which can be used in the laboratory
2) Easily create accurate scale down models for process development, membrane screening and process
characterization at the micro-scale
3) The fluid path is drainable and designed for Clean-in-Place(CIP) using industry standard cleaning agents.
4) Tank material is Acrylic see in Product
5) MF/UF membrane can be used
6) Linear scale-up
7) Robust design with minimal maintenance requirement
8) Possible minimum working volume : 15 mL
9) Stir base system to mix solution in the reservoir
10) This system is able to control TMP using pressure control of circulation flow
11) User's manual
3. Feature of Membrane
1) Maximum Transmembrane Pressure : 5 bar
2) Promotor in the membrane channel(from A screen to V screen): to generate turbulence flow
3) Polyurethane Frame to keep linear path length
4) Low protein binding MF membrane material(PVDF) of module
5) 2 kinds type of UF membrane of UF membrane module
6) Void Free UF membrane
7) Validation Guide/Maintenance Guide
Part II
1. Applications
1) Concentration and diafiltration
2) Depyrogenation
3) Desalting and buffer exchange
4) Small molecule processing
5) Cell harvesting/ clarification
6) Virus harvesting/ clarification
2. Feature of System
1) Load cell
2) Automatic TMP(Trans Membrane Pressure) control suited for process development
3) Tank material is Epoxy or Borosilicate Glass that see in Product
4) sanitary-design according to cGMP concept
5) MF/UF membrane can be used
6) Linear scale-up
7) Closed special tank-in-tank design
8) Possible minimum working volume : 170- 300 mL
9) System is designed for cGMP batch or scale up/scale down studies.
10) Design through combination of minimal pumping and a special vertical orientation of pump
(diaphragm pump)
3. Feature of Membrane
1) Maximum Transmembrane Pressure : 4 - 6bar
2) Polyurethane Frame to keep linear path length
3) Low protein binding MF membrane material(PVDF) of UF membrane module
4) Void Free UF membrane
5) Wide range of membrane material, cut-off size(5kD~1000kD), pore-size(0.1um~0.65um)
6) Validation Guide/Maintenance Guide
B. SPECIFICATION
Part I
1. Cassette
UF membrane (not included) ( filter area : 150 cm2)
2. Cassette Holder
UF membrane holder
3. Vent Filter
Air Filter
4. Filtration Area
50 cm2 – 150 cm2
5. Tank Volume & Meterial
500 ml, Acrylic tank ( Product see proccessing )
6. Hold-up Volume
< 3.2 mL (excluding cassette)
7. Pumps
Feed ( Peristaltic pump ) : 10 – 100 ml /min
8. Maximum System Pressure
5.5 barg
9. Minimum Recirculation Volume at Maximum Flow Rate
15 mL
10. Process Temperature
4 – 45 °C ( It can use in cold roon )
11. Operational Requirements
Electrical : Voltage/Current — Single Phase
230 VAC 50 Hz
Tubing : LFL Tygon tubing
Luer fittings : 316 stainless steel/Polypropylene
Reservoir cover : Polypropylene
Retentate valve diaphragm : EPDM and polypropylene
Stir bar : PTFE
Housing/enclosure : Polyurethane
12. Environmental Operating Conditions
Humidity : 10 – 85% non-condensing environment
Temperature range : Ambient operational: 4 °C to 30 °C
Process: 4 °C to 50 °C
Storage: –10 °C to 55 °C
Maximum altitude : < 2000 meters
13. Dimensions
Height: 45.7 cm (18 in.)
Width: 21.6 cm (8.5 in.)
Depth: 27.9 cm (11 in.)
Weight: 5.9 kg (13 lbs)
14. Wetted Materials
All polymer wetted materials of construction are USP Class VI tested.
15. System Setup
1) Unit shipped requires setup.
2) UF membrane (not included) are installed by supplier & user. ( in Basic Trainning )
3) Operating is by supplier & user ( in Basic Trainning )
16. Accessary
1) UFmembrane 50cm2 12pk
2) Multi Manifolder 4pk
3) One Way Stop Cock Valve 1pk
4) Tank vent filter 1pk
Part II
1. Cassette
UF membrane(not included) ( filter area : 0.1 - 0.5m2 )
2. Cassette Holder
UF membrane holder
3. Vent Filter
Air filter membrane
4. Filtration Area
0.1 m2 – 2.0 m2
5. Tank Volume & Meterial
5 ~ 10 L, Tank meterial is Epoxy or Borosilicate Glass that see in Product
6. Tank Mixer Volume
1.5 - 10 L
7. Hold-up Volume
5 ~ 10 mL (excluding cassette)
8. Pumps
Feed ( Diaphram pump ) : 0.2 – 12 L/min at 6 barg
Transfer ( Peristaltic pump ) (option): 0.16 – 2300 mL/min (tubing dependent)
9. Heat Exchanger (option)
0.2 kW, 48kcal control
10. Retentate Flow Meter (option)
Minimum conductivity: ≥ 5 µS/cm
Accuracy: ± 0.5%
11. Maximum System Pressure
6 barg prior to pump shut off
12. Minimum Recirculation Volume
170ml - 300 mL
13. Process Temperature
0 – 50 °C ( It can use in cold roon )
14. Connection Hoses
1.0 bar maximum
15. Operator Interface
PLC via a touch screen
16. Operational Requirements
Electrical : Voltage/Current — Single Phase
100 VAC 50 Hz, 100 VAC 60 Hz,
120 VAC 60 Hz, 230 VAC 50 Hz
Amperage (maximum) : Europe: 2.5 amps
North America and Japan: 5 amps
Over voltage category : Category II
Compressed air : Maximum 60 psi/413 kPa
–20 °C dew point, oil-free
17. Environmental Operating Conditions
Humidity : 10 – 85% non-condensing environment
Temperature range : Ambient operational: 2 °C to 30 °C
Process: 0 °C to 50 °C
Storage: –10 °C to 55 °C
Maximum altitude : < 2000 meters
18. Dimensions
Height: 94 - 110 cm (37 - 45 in.)
Width: 48 - 80 cm (19 - 32 in.)
Depth: 44 - 68 cm (18 - 27 in.)
Weight: 50 - 59 kg
19. Wetted Materials of Construction
EPDM, 316L stainless steel, Santoprene® elastomer, medical grade epoxy, and silicone
20. Surfaces
Stainless steel – internal : < 0.6 - 0.8 µm, Ra
External : < 0.8 - 1.34 µm, Ra
21. Wetted Materials
All polymer wetted materials of construction are USP Class VI tested.
22. System Setup
1) Unit shipped requires setup.
2) UF membrane (not included) are installed by supplier & user. ( in Basic Trainning )
3) Operating is by supplier & user ( in Basic Trainning )
23. Regulatory Information
The UF system meets the requirements of the Low Voltage Directive
73/23/EEC and the EMC Directive 89/336/EEC, and is CE marked.
24. Accessary
1) UF membrane 0.1m2 3pk
2) Retentate Flow Meter kit 1pk
3) Transfer pump console(230VAC) 1pk
IV. Remarks
1. The installation and operation should be performed by bidder at user side.
2. Two years warranty service should be provided by supplier after the date of installation.
3. After service certification should be submitted by local agency